Steris Corporation recalls VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Products
  • VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
10724995220331
Units affected
409
About this recall
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007