Steris Corporation recalls
Steris Corporation is named on 12 recalls in Recall Relay's records since November 2023, from FDA (26 recalled products: FDA lists each product as its own recall); the most recent was September 23, 2026.
All 26 recalled products, newest first.
-
Class IIofficial
Steris Corporation recalls HexaVue Focus Integration SystemA software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation discrepancy or incorrect association of clinical images with a patient record.
-
Class IIofficial
Steris Corporation recalls RAPICIDE PA Part A High-Level DisinfectantIf the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
-
Class IIofficial
Steris Corporation recalls RAPICIDE PA Part A High-Level DisinfectantIf the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
-
Class IIofficial
Steris Corporation recalls RAPICIDE PA Part A High-Level DisinfectantIf the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
-
Class IIofficial
Steris Corporation recalls VERIFY Biological Steam Test Pack Model/Catalog Number: S3069Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
Class IIofficial
Steris Corporation recalls VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
Class IIofficial
Steris Corporation recalls VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
Class IIofficial
Steris Corporation recalls VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
Class IIofficial
Steris Corporation recalls VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
Class IIofficial
Steris Corporation recalls VERIFY Biological Steam Test Pack Model/Catalog Number: S3065Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
Class IIofficial
Steris Corporation recalls ALYON Surgical Lighting SystemThe the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
-
Class IIofficial
Steris Corporation recalls HarmonyAIR A-Series Surgical Lighting SystemThere is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.
-
Class IIofficial
Steris Corporation recalls HarmonyAIR A-Series Surgical Lighting SystemThe the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
-
Class IIofficial
Steris Corporation recalls AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/DisiWire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
-
Class IIofficial
Steris Corporation recalls AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/DisiWire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
-
Class IIofficial
Steris Corporation recalls Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting SystemsThe DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
-
Class IIofficial
Steris Corporation recalls Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting SystemsThe DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
-
Class IIofficial
Steris Corporation recalls Barco MNA with the HexaVue IP Integration SystemThis is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.
-
Class IIofficial
Steris Corporation recalls Hookup Model/Catalog Number: DSD-110-HU0163 Product Description: Hookups provide the interface between DSD AutIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
-
Class IIofficial
Steris Corporation recalls Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an AdvantagIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
-
Class IIofficial
Steris Corporation recalls Hookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an Advantage PIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
-
Class IIofficial
Steris Corporation recalls Hookup Model/Catalog Number: 2-8-540CAS Product Description: Hookups provide the interface between an AdvantagIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
-
Class IIofficial
Steris Corporation recalls Dual Articulating HeadrestDual-articulating head rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay.
-
Class IIofficial
Steris Corporation recalls Dual Articulating HeadrestDual-articulating head rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay.
-
Class IIofficial
Steris Corporation recalls V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-DutyBandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.
-
Class IIofficial
Steris Corporation recalls Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope ReprocIdentification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.