Steris Corporation recalls RAPICIDE PA Part A High-Level Disinfectant
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
- Products
-
- RAPICIDE PA Part A High-Level Disinfectant
- Company
- Steris Corporation
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00724995245962- Units affected
- 106758 units
- About this recall
- RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117