Steris Corporation recalls VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
- Products
-
- VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061
- Company
- Steris Corporation
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
50724995023824- Units affected
- 2205
- About this recall
- VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061