Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Steris Corporation recalls VERIFY Biological Steam Test Pack Model/Catalog Number: S3069

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Products
  • VERIFY Biological Steam Test Pack Model/Catalog Number: S3069
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
10724995023864
Units affected
233
About this recall
VERIFY Biological Steam Test Pack Model/Catalog Number: S3069