Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1100

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Products
  • Leica Biosystems Cryostat Model Number CM1100
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
04049188066237
Units affected
1 unit
About this recall
Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples