Leica Biosystems Nussloch GmbH recalls
Leica Biosystems Nussloch GmbH is named on 5 recalls in Recall Relay's records since November 2023, from FDA (10 recalled products: FDA lists each product as its own recall); the most recent was January 30, 2025.
All 10 recalled products, newest first.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software VersionFirm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1950As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1510 SAs part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM3050 SAs part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1850As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1100As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1900As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica HistoCore Arcadia HAn issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica HistoCore PEGASUS Tissue ProcessorPoorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.
-
Class IIofficial
Leica Biosystems Nussloch Gmbh recalls Leica HistoCore PEGASUS (PROCESSORThere is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.