Leica Biosystems Nussloch Gmbh recalls Leica Biosystems Cryostat Model Number CM1510 S
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
- Products
-
- Leica Biosystems Cryostat Model Number CM1510 S
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
0404918800294504049188002891- Units affected
- 13 units
- About this recall
- Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples