Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

American Contract Systems Inc. recalls Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Products
  • Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR
Where it was sold

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States: Iowa, Illinois, Minnesota, South Dakota, Washington

Barcodes
191072214362191072235169191072212474191072237446191072235299191072216434191072200495191072150431191072229052191072235602
Units affected
32,433 total
About this recall
Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA75I VASCULAR PACK