Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

American Contract Systems Inc. recalls Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) U

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Products
  • Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) U
Where it was sold

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States: Iowa, Illinois, Minnesota, South Dakota, Washington

Barcodes
19107223436019107220933719107223655519107223267019107210667400191072234117
Units affected
32,433 total
About this recall
Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) UIHY16AE ASC HYSTEROSCOPY PACK (PS 020634)