American Contract Systems Inc. recalls Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CME
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Products
-
- Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CME
- Where it was sold
-
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
States: Iowa, Illinois, Minnesota, South Dakota, Washington
- Barcodes
191072225955191072211514191072199720191072202918191072224569- Units affected
- 32,433 total
- About this recall