Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual indicator (exposure LED light) the operator can monitor the x ray release. Also, after a successful x ray exposure, the clinical image is displayed on the monitor.
Products
  • Mobilett Mira wireless (VA20) mobile x-ray system(s)
Where it was sold

US Nationwide - Worldwide

Nationwide

Barcodes
04056869009117
Units affected
4 distributed in U.S.
About this recall