Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40A
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
- Products
-
- ARTIS Pheno VE30A and VE40A
- Where it was sold
-
U.S. and O.U.S.
- Barcodes
04056869046877- Units affected
- 73
- About this recall
- ARTIS Pheno VE30A and VE40A, Model 10849000