Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Aizu Olympus Co., Ltd. recalls Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Products
  • Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
823 units
About this recall
Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.