Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Aizu Olympus Co., Ltd. recalls GIF-H190

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Products
  • GIF-H190
Where it was sold

Domestic distribution to FL, KY, MA, NJ, NY, and PA

States: Florida, Kentucky, Massachusetts, New Jersey, New York, Pennsylvania

Barcodes
04953170305290
Units affected
24 incorrectly repaired units
About this recall
GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE