Aizu Olympus Co., Ltd. recalls
Aizu Olympus Co., Ltd. is named on 3 recalls in Recall Relay's records since November 2023, from FDA (30 recalled products: FDA lists each product as its own recall); the most recent was October 31, 2025.
All 30 recalled products, newest first.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Olympus OER-MiniInform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Olympus OER-ProInform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Olympus OER-EliteInform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls GIF-HQ190Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls GIF-H190Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls PCF-H190DLSpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls PCF-H190LSpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Mobile airway scope-diagnosis and observation to access airway anatomyChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Ultrasonic gastrovideoscope-Real time ultrasound imagingChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anusChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Small intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including tChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps)Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagusChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Tracheal videoscope- For airway managementChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Cystonephroscope-endoscopic diagnosis and treatment within the bladderChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model NumChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Ureteralscope- for endoscopic diagnosis and treatment within the ureterChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imagingChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Colonoscope-indicated for use within the lower digestive tract (including the anusChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (includingChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers:Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIIofficial
Aizu Olympus Co., Ltd. recalls Choleodochoscope. Model Numbers: CHF-BP30Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Gastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery withinChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Tracheal fiberscope-airway managementChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anusChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.