Aizu Olympus Co., Ltd. recalls Choleodochoscope. Model Numbers: CHF-BP30
FDA Class III
FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).
- Hazard
- Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- Products
-
- Choleodochoscope. Model Numbers: CHF-BP30
- Company
- Aizu Olympus Co., Ltd.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 3 units
- About this recall
- Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.