Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Aizu Olympus Co., Ltd. recalls Choleodochoscope. Model Numbers: CHF-BP30

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Products
  • Choleodochoscope. Model Numbers: CHF-BP30
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
3 units
About this recall
Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.