Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Aizu Olympus Co., Ltd. recalls Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Products
  • Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
46 units
About this recall
Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-VT3.