Aizu Olympus Co., Ltd. recalls Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- Products
-
- Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus
- Company
- Aizu Olympus Co., Ltd.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 8 units
- About this recall
- Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.