Aizu Olympus Co., Ltd. recalls Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Num
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- Products
-
- Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Num
- Company
- Aizu Olympus Co., Ltd.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 25 units
- About this recall
- Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.