Aizu Olympus Co., Ltd. recalls Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- Products
-
- Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps)
- Company
- Aizu Olympus Co., Ltd.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 0 units
- About this recall
- Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.