Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Shimadzu Corporation Medical Systems Division recalls Trinias Opera Smart (C16s

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
Products
  • Trinias Opera Smart (C16s
Where it was sold

US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.

States: California, Florida, New York, South Carolina, Tennessee, Texas

Units affected
118 unts.
About this recall
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),