Aizu Olympus Co., Ltd. recalls GIF-HQ190
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
- Products
-
- GIF-HQ190
- Company
- Aizu Olympus Co., Ltd.
- Where it was sold
-
Domestic distribution to FL, KY, MA, NJ, NY, and PA
States: Florida, Kentucky, Massachusetts, New Jersey, New York, Pennsylvania
- Barcodes
04953170305276- Units affected
- 24 incorrectly repaired units
- About this recall
- GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE