Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade. Model Number: 11689304.
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Products
-
- MAGNETOM Cima.X Upgrade. Model Number: 11689304.
- Where it was sold
-
US distribution to California. International distribution to Australia, Germany, Great Britian.
States: California
- Barcodes
04056869975245- Units affected
- 4 system (1 US; 3 OUS)
- About this recall
- MAGNETOM Cima.X Upgrade. Model Number: 11689304.