Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade. Model Number: 11689304.

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Products
  • MAGNETOM Cima.X Upgrade. Model Number: 11689304.
Where it was sold

US distribution to California. International distribution to Australia, Germany, Great Britian.

States: California

Barcodes
04056869975245
Units affected
4 system (1 US; 3 OUS)
About this recall
MAGNETOM Cima.X Upgrade. Model Number: 11689304.