Medtronic Perfusion Systems recalls MC3 VitalFlow Console

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
Products
  • MC3 VitalFlow Console
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
10850007732247
Units affected
167 units
About this recall
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo