FDAMedical devicesRecalled Status: Ongoing

GE Healthcare (China) Co., Ltd. recalls GE Healthcare Operator Console Upgrade Types

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Products
  • GE Healthcare Operator Console Upgrade Types
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
00195278505798
Units affected
4 units
About this recall
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (for BrightSpeed, Elite/BrightSpeed Elite Select), Non-EU: BrightSpeed Elite, BrightSpeed Elite Select; X-ray/computed tomography system