Abiomed, Inc. recalls Oscor Intro Kit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
- Products
-
- Oscor Intro Kit
- Company
- Abiomed, Inc.
- Where it was sold
-
Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK
Nationwide
- Units affected
- 1 unit within kit
- About this recall
- Oscor Intro Kit, 14Fr x 13 and 25 cm Part of Pump Set (0048-0040)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3025