Abiomed, Inc. recalls
Abiomed, Inc. is named on 26 recalls in Recall Relay's records since November 2023, from FDA (75 recalled products: FDA lists each product as its own recall); the most recent was September 18, 2026.
All 75 recalled products, newest first.
-
Class Iofficial
Abiomed, Inc. recalls Impella Connect Optical AICRetrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
-
Class Iofficial
Abiomed, Inc. recalls Impella Optical ControllerRetrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella ControllerRetrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
-
Class Iofficial
Abiomed, Inc. recalls Impella Optical ControllerThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class Iofficial
Abiomed, Inc. recalls Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella OptThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. ImpThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Introducer Kit 23 FrA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP Smart Assist SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Intro KitA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella 5.5 Smart Assist SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella RP Flex with SmartAssist SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella 5.5 SmartAssist S2 SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP Introducer Kit 14 FrA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Insertion KitA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP Introducer Kit 14 FrA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP Introducer Kit 14 FrA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls ABIOMED Impella CP SetA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
-
Class Iofficial
Abiomed, Inc. recalls Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer KitsPotential for thrombus formation during prolonged use of the introducer.
-
Class Iofficial
Abiomed, Inc. recalls Abiomed 14 Fr x 13 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000434. IntroducerPotential for thrombus formation during prolonged use of the introducer.
-
Class Iofficial
Abiomed, Inc. recalls Abiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000435. IntroducerPotential for thrombus formation during prolonged use of the introducer.
-
Class Iofficial
Abiomed, Inc. recalls Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer KitsPotential for thrombus formation during prolonged use of the introducer.
-
Class Iofficial
Abiomed, Inc. recalls Impella CP Set with SmartAssist. Product Codes: 1000080Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC); Product Code: 0042-0000-US;A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.
-
Class Iofficial
Abiomed, Inc. recalls Purge CassetteIncreased risk of purge leaks with Generation 1 purge cassettes.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
-
Class Iofficial
Abiomed, Inc. recalls Impella RP. Product Code: 0046-0011.Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
-
Class Iofficial
Abiomed, Inc. recalls Impella RP with SmartAssist. Product Code: 0046-0035.Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
-
Class Iofficial
Abiomed, Inc. recalls Impella RP Flex with SmartAssist. Product Code: 1000323.Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 with SmartAssist¿ S2Device packaged in incorrect outer box carton.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC) labeled as the following with corresponding Product Codes: 1. AIC w/ImpellaPotential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. ImSpecific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
-
Class Iofficial
Abiomed, Inc. recalls Impella RP with SmartAssist; Product Number: 0046-0035;Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
-
Class Iofficial
Abiomed, Inc. recalls Impella RP Flex with SmartAssist; Product Number: 1000323;Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 with SmartAssist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 Set AU; Product Code: 1000361;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Pump 371 14F LT CMR Set; Product Code: 0048-0045;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Pump SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Smart Assist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 with SmartAssist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 with SmartAssist S2 SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Smart Assist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Smart Assist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 with SmartAssist S2 SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 SmartAssist S2 SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Smart Assist Set; Product Code: 1000402;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Smart Assist Set APAC; Product Code: 0048-0047;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 with SmartAssist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Pump 381 Pump Set (US); Product Code: 1000080;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella 5.5 SmartAssist SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls Impella CP Pump 371 SetA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class IIofficial
Abiomed, Inc. recalls ImpellaCP SmartAssistSetAPACA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
-
Class Iofficial
Abiomed, Inc. recalls Impella CP with SmartAssistNine (9) Impella CP pumps failed inspection and were inadvertently released.
-
Class Iofficial
Abiomed, Inc. recalls Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) ProIFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
-
Class IIofficial
Abiomed, Inc. recalls Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) ProNew warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.
-
Class IIofficial
Abiomed, Inc. recalls Impella ConnectThe following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature; Sort case tiles by AIC alarm color feature; Pump metric display on case tile feature
-
Class IIofficial
Abiomed, Inc. recalls Abiomed KITHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Abiomed Introducer KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis.
-
Class IIofficial
Abiomed, Inc. recalls Abiomed KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Oscor Introducer KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Abiomed KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Abiomed Oscor Introducer KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Abiomed KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Abiomed Axillary Insertion Introducer -Introducer(s) are intended for introduction of the Impella Catheter intHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Oscor Intro KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Abiomed KitHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Pump not detected as connected to controller due to software issue.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Pump not detected as connected to controller due to software issue.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. ImpRetrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. ImpRetrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.