Abiomed, Inc. recalls Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Opt
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
- Products
-
- Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Opt
- Company
- Abiomed, Inc.
- Where it was sold
-
Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;
Nationwide
- Barcodes
0081350201093000813502010961008135020114250081350201097800813502010985- Units affected
- 1422 units (487 US, 935 OUS)
- About this recall
- Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 3. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 4. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 5. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US.