CooperSurgical, Inc. recalls Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Prod
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
- Products
-
- Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Prod
- Company
- CooperSurgical, Inc.
- Where it was sold
-
States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, MA, MK, MT, NL, NO, PE, PH, PK, PT, QA, RO, RS, SA, SD, SE, SO,TR, TW, TZ, US, ZA
States: California, Illinois, North Carolina, New Jersey, New York, Pennsylvania, Texas, Virginia
- Units affected
- 37 units
- About this recall
- Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: No