Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Healthcare Diagnostics, Inc. recalls Magnesium Flex reagent cartridge

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out
Products
  • Magnesium Flex reagent cartridge
Where it was sold

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Slovenia, Spain.

Nationwide

Barcodes
00842768015694
Units affected
13,115 units (11,520 US, 1,595 ROW)
About this recall
Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;