Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Healthcare Diagnostics, Inc. recalls Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
Products
  • Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL
Where it was sold

Worldwide - US Nationwide distribution in the state of CA and the country of Italy.

States: California

Barcodes
00842768001048
Units affected
20 units
About this recall
Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL