Siemens Healthcare Diagnostics, Inc. recalls Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
- Products
-
- Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL
- Where it was sold
-
Worldwide - US Nationwide distribution in the state of CA and the country of Italy.
States: California
- Barcodes
00842768001048- Units affected
- 20 units
- About this recall
- Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL