Abiomed, Inc. recalls Impella RP Flex with SmartAssist; Product Number: 1000323;
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
- Products
-
- Impella RP Flex with SmartAssist; Product Number: 1000323;
- Company
- Abiomed, Inc.
- Where it was sold
-
Domestic: Nationwide Distribution International: Canada
Nationwide
- Barcodes
00813502012811- Units affected
- 2,364 units
- About this recall
- Impella RP Flex with SmartAssist; Product Number: 1000323;