Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Medical Solutions USA, Inc recalls MAMMOMAT Revelation;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
Products
  • MAMMOMAT Revelation;
Where it was sold

Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of Australia, Canada, India.

States: Arkansas, California, New York

Barcodes
04056869124575
Units affected
7 units (3 US, 4 OUS)
About this recall
MAMMOMAT Revelation;