Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Medical Solutions USA, Inc recalls LUMINOS Lotus Max

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.
Products
  • LUMINOS Lotus Max
Where it was sold

U.S. and OUS

Units affected
U.S.: 1, OUS: 95
About this recall
LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471