Siemens Medical Solutions USA, Inc recalls Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Sof

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
Products
  • Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Sof
Where it was sold

Worldwide - US Nationwide distribution in the states of AK, CA, IL, MN, NY, OR, VA, WI and the countries of Austria, Belgium, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland.

States: Alaska, California, Illinois, Minnesota, New York, Oregon, Virginia, Wisconsin

Barcodes
04056869993973
Units affected
41 units
About this recall
Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02