Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Erythropoietin (EPO) Assay (100 Test)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
Products
  • Atellica IM Erythropoietin (EPO) Assay (100 Test)
Where it was sold

Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, U.A.E., United Kingdom.

States: Alabama, California, Iowa, Illinois, Michigan, Minnesota, North Carolina, Nebraska, Nevada, New York, Ohio, Texas

Barcodes
00630414245775
Units affected
1,057 units (46 US, 1,011 OUS)
About this recall
Atellica IM Erythropoietin (EPO) Assay (100 Test)