Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
- Products
-
- ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, U.A.E., United Kingdom.
States: Alabama, California, Iowa, Illinois, Michigan, Minnesota, North Carolina, Nebraska, Nevada, New York, Ohio, Texas
- Barcodes
00630414010380- Units affected
- 264 units (6 US, 258 OUS)
- About this recall
- ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)