Siemens Healthcare Diagnostics, Inc. recalls DCA Systems HbA1c Reagent Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
Products
  • DCA Systems HbA1c Reagent Kit
Where it was sold

Worldwide distribution - US Nationwide and the country of Taiwan.

Nationwide

Barcodes
00630414532806
Units affected
74,211 Kits (Expanded 2/28/24: additional 72,892 kits)
About this recall
DCA Systems HbA1c Reagent Kit