Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Products
  • IBA Proton Therapy System - PROTEUS 235
Where it was sold

Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.

States: Florida

Barcodes
05404013801138
Units affected
4 units (1 US, 3 OUS)
About this recall
IBA Proton Therapy System - PROTEUS 235