Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
- Products
-
- IBA Proton Therapy System - PROTEUS 235
- Company
- Ion Beam Applications S.A.
- Where it was sold
-
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
States: Florida
- Barcodes
05404013801138- Units affected
- 4 units (1 US, 3 OUS)
- About this recall
- IBA Proton Therapy System - PROTEUS 235