Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of p

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
Products
  • IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of p
Where it was sold

Worldwide - US Nationwide distribution in the state of Virginia and the country of India.

States: Virginia

Units affected
2 units
About this recall
IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.