GE Medical Systems China Co., Ltd. recalls GE Healthcare CARESTATION 650C A1

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Products
  • GE Healthcare CARESTATION 650C A1
Where it was sold

Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.

Nationwide

Barcodes
00840682103954
Units affected
69 units
About this recall
GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.