Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Medical Solutions USA, Inc recalls ARTIS pheno and ARTIS Icono biplane

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Products
  • ARTIS pheno and ARTIS Icono biplane
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
4056869046877405686906331740568692959234056869149325
Units affected
U.S. 758, OUS 2077
About this recall