Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.
Products
  • Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
Where it was sold

Domestic: Nationwide Distribution International: Argentina, Australia, Bangladesh, Brazil, Bulgaria , Canada, Chile, China, Colombia, Croatia , Cura¿ao, St Eus, Czech Republic , Denmark , Ecuador, Egypt , Estonia , Finland , France , Georgia , Germany , Greece , Hong Kong, India, Iraq , Italy , Jordan , Kuwait , Latvia , Lithuania , Macedonia , Malaysia, Maldives, Mexico, Netherlands , New Zealand, Norway , Oman , Pakistan , Paraguay, Qatar , Romania , Saudi Arabia , Serbia , Singapore, Slovakia , Slovenia , South Africa, South Korea, Spain , Turkey , Turkmenistan , U.A.E. , Uganda , United Kingdom , Uruguay , Vatikancity;

Nationwide

Barcodes
00630414220505
Units affected
21,291 units (9750 US, 11541 OUS)
About this recall
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;