Belmont Instrument LLC recalls Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
- Products
-
- Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile
- Company
- Belmont Instrument LLC
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.
States: Arizona, California, Colorado, Connecticut, District of Columbia, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Maine, Minnesota, North Carolina, North Dakota, New Jersey, New Mexico, Ohio, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia
- Barcodes
00896128002213- Units affected
- 208 sets
- About this recall
- Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).