Belmont Instrument LLC recalls 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
- Products
-
- 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume
- Company
- Belmont Instrument LLC
- Where it was sold
-
Nationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI. International distribution to Czechia, Italy, Norway.
Nationwide
- Barcodes
00896128002534- Units affected
- 603 cases (2,412 sets)
- About this recall
- 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.