FDAMedical devicesRecalled Status: Ongoing

Belmont Instrument LLC recalls Allon 2001. Part Number: 200-00263

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.
Products
  • Allon 2001. Part Number: 200-00263
Where it was sold

Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.

States: California, District of Columbia, Louisiana, Massachusetts, New Mexico, Pennsylvania, Texas, Virginia, Washington

Units affected
229 units
About this recall
Allon 2001. Part Number: 200-00263