Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
- Products
-
- Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
Nationwide
- Barcodes
00630414598253- Units affected
- 2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS
- About this recall
- Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498