Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
- Products
-
- ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
Nationwide
- Barcodes
00630414468198- Units affected
- 1,072 packs (2,144 units) - 156 packs US, 916 packs OUS
- About this recall
- ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.