Hamilton Medical AG recalls Ventilator HAMILTON-C6

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Products
  • Ventilator HAMILTON-C6
Where it was sold

US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.

States: California, Florida, Georgia, Indiana, Kentucky, Michigan, Nevada, Oregon, Pennsylvania, South Carolina, Texas, Virginia

Barcodes
07630002808590
Units affected
68
About this recall
Ventilator HAMILTON-C6, PN: 160021